Contact Us Cart Site Map Manuals Order Products Home

3 1/2 Digit Thermocouple Meters
MODEL 202A-J, -K, -T

      
Electromagnetic Compatibility

NEWPORT PRODUCT INFO
MANUAL - PDF Version
MECHANICAL 1/8 DIN (2A) Case
MECHANICAL Connections Diagram
PRICE
REQUIRES ADOBE ACROBAT - HELP
STANDARD FEATURES
J, K or T thermocouple input
1° or 0.1° resolution
°C or °F, jumper-selectable
0.05 deg/deg cold-junction error
Five-segment linearization
1 mV/deg linearized recorder output
120 dB CMR, 70 dB NMR
Bright, 0.56 in (14.2 mm) LED display
Display hold and test
115/230 Vac power
EMI/RFI filter for AC power
Screw-terminal barrier strip
Short (4.1 in deep) 1/8 DIN case
 
OPTIONS
Isolated 9-32 V dc power or 26-56 Vdc power
NEMA-4 splash-proof lens cover

Models 202A-J, -K and -T
are 3 1/2 digit ±1,999 count) panel meters for thermocouple types J, K and T. They are available for display in °C or °F and resolution of 1° up to 1,999° or 0.1° up to 199.9°. They are recommended for demanding indication-only applications where exceptionally low-cost is required. For display up to 1,999.9° with 0.1° resolution, Newport recommends its 2002A Series meters, which provides a 4 1/2 digit (±19,999 count) main assembly.

STANDARD FEATURES
Newport's 202A Series thermocouple meters provide exceptional accuracy, as made possible by five-segment linearization and use of matched, graded components for each T/C type. Electrical features include cold-junction compensation and jumper selection of °C or °F. A 1 mV per count linearized analog output is standard and is ideal for driving a strip-chart recorder.

Mechanical features include a bright 0.56 in (14.2 mm) display, a screw-terminal barrier strip for signal and power, and a compact 1/8 DIN case that requires less than 4.1 in (104 mm) behind the panel.

STANDARD FEATURES
Mechanical features include a bright 0.56 in (14.2 mm) display, a screw-terminal barrier strip for signal and power, and a compact 1/8 DIN case that requires less than 4.1 in (104 mm) behind the panel. Electrical features include cold-junction compensation, jumper selection of °C or °F, and a linearized recorder output of 1 mV per degree.

OPTIONS
Options are isolated 9-32 V dc or 26-56 V dc power, a PCB edge connector for display control and output of +4.7 Vdc and ­4.7 Vdc, and a splash-proof lens cover which meets NEMA-4 requirements.




SPECIFICATIONS

INPUT
Thermocouple: Type J (Iron-Constantan, Fe-CuNi)
Type K (Chromel-Alumel, NiCr-NiAl)
Type T (Copper-Constantan, Cu-CuNi)
Calibration: IEC 584-1 (IPTS-68)
Configuration: Single-ended (-TC lead connected to ANA GND)
Polarity: Bipolar
Zero: Adjustable ±20°C (±36°F)
Cold-junction tempco: ±0.05 deg/deg

Sensor-wire resistance effect per conductor:
Type J: 40 µdeg/deg/ohm, up to 2500 ohm
Type K: 50 µdeg/deg/ohm, up to 2000 ohm
Type T: 50 µdeg/deg/ohm, up to 2000 ohm
Sensor-break current: 300 nA
Sensor-break indication: Meter displays positive overrange (upscale)


NOISE REJECTION
NMR, SIG HI to SIG LO:
70 dB, 50/60 Hz
CMR, ANA GND to PWR GND: 120 dB, DC to 60 Hz
CMV, ANA GND to PWR GND: 1500 Vp per HV test, 354 Vp per IEC spacing

ACCURACY AT 25°C
(1° resolution models)
Range for rated accuracy:
Type J: ­138 to +760°C (­216 to +1400°F)
Type K: ­105 to +1248°C (­157 to +1999°F)
Type T: ­112 to +400 °C (­170 to +752°F)
Maximum error: ±1.5°C (±2.7°F) ±1/2 count
Span tempco: ±0.01% of reading/°C
Resolution: 0.1°C or 0.1°F (jumper-selectable)
Full-scale step response: 1 s
Warmup to rated accuracy: 10 min

ACCURACY AT 25°C
(0.1° resolution models)
Range for rated accuracy, all T/C types: ­105.0 to 199.9°C (­157.0 to 199.9°F)
Maximum error: ±1.0°C (±1.8°F) ±1/2 count
Span tempco: ±0.01% of reading/°C
Resolution: 0.1°C or 0.1°F (jumper-selectable)
Full-scale step response: 1 s
Warmup to rated accuracy: 20 min


ANALOG OUTPUT (linearized)
Voltage: 1 mV/count
Calibration error: ±1.5 mV on °C, ±2.7 mV on °F
Current: 1 mA max
Source resistance: 68 ohm


ANALOG-TO-DIGITAL CONVERSION
Technique:
Dual-slope, average value
Signal-integration period: 100 ms
Read rate: 2.5/s

DISPLAY
Type:
7-segment, red LED
Height: 0.56 in (14.2 mm)
Symbols: ­1.8.8.8
Overrange indication: Three least-significant digits blank

DIGITAL INPUTS
Level: TTL or 5 V CMOS compatible

POWER

AC voltages:
115 or 230 V ac, ±15%
AC frequency: 49-440 Hz
DC voltages: 9-32 V dc, isolated to 300 Vp; 26-56 V dc, isolated to 300 Vp
Power consumption: 3.7 W
Output voltages: +4.7 V dc and ­4.7 V dc ±5%, 10 mA max

ENVIRONMENTAL
Operating temperature: 0 to 60°C
Storage temperature: ­-40 to +85°C
Relative humidity: 95% at 40°C (non-condensing)

MECHANICAL
Dimensions:
Newport DIN2A (short 1/8 DIN case) (see Mechanical section for drawings)
Weight: 15 oz (425 g)
Case material: 94V-0 UL-rated polycarbonate



To Order (Specify Model Number, see variations below in partnumber builder table)    
 Part Number   Availability   Price    Description    Qty.
 Thermocouple Input, Type J, Iron - Constantan (Fe-CuNi)
 202A-JC1 in-stock    $285.00     Thermocouple meter, J thermocouple, -138 to +760°C temperature range  
 202A-JC2 in-stock    $285.00     Thermocouple meter, J thermocouple, -105.0 to +199.9°C temperature range  
 202A-JF1 in-stock    $285.00     Thermocouple meter, J thermocouple, -216 to +1400°F temperature range  
 202A-JF2 in-stock    $285.00     Thermocouple meter, J thermocouple, -157.0 to +199.9°F temperature range  
 Thermocouple Input, Type K, Chromel-Alumel (NiCr-NiAl)
 202A-KC1 in-stock    $285.00     Thermocouple meter, K thermocouple, -105 to +1248°C temperature range  
 202A-KC2 in-stock    $285.00     Thermocouple meter, K thermocouple,-105.0 to +199.9°C temperature range  
 202A-KF1 in-stock    $285.00     Thermocouple meter, K thermocouple, -157 to +1999°F temperature range  
 202A-KF2 in-stock    $285.00     Thermocouple meter, K thermocouple,-157.0 to +199.9°F temperature range  
 Thermocouple Input, Type T, Copper-Constantan (Cu-CuNi)
 202A-TC1 in-stock    $285.00     Thermocouple meter, T thermocouple,-112 to +400°°C temperature range  
 202A-TC2 in-stock    $285.00     Thermocouple meter, T thermocouple,-105.0 to +199.9°°C temperature range  
 202A-TF1 in-stock    $285.00     Thermocouple meter, T thermocouple,-170 to +752°F temperature range  
 202A-TF2 in-stock    $285.00     Thermocouple meter, T thermocouple,-157.0 to +199.9°F temperature range  
 
All dollar amounts on this site are shown in US currency.

 Part Number Builder
(1) (2) (3) (4) (5)
Option Descriptions
(1) Models
  Select
202A for standard
302 for model with INFINITY® style bezel and adapter
(2) Thermocouple Input Configurations
  Select
-JC1 for -138 to +760°C. (JB1)
-JC2 for -150.0 to +199.9°C (JB2)
-JF1 for -216 to +1400°F (JB1)
-JF2 for -157.0 to +199.9°F
-KC1 for -105 to +1248°C (KB1)
-KC2 for -105.0 to +199.9°C (KB2)
-KF1 for -157 to +1999°F (KB1)
-KF2 for -157.0 to +199.9°F (KB2)
-TC1 for -122 to +400°C (TB1)
-TC2 for -105.0 to +199.9°C (TB2)
-TF1 for -170 to +752°F (TB1)
-TF2 for -157.0 to +199.9°F (TB2)
(3) POWER
  Select
Nothing(leave field blank) for 115 Vac, 50/60 Hz
,C1 for 230 Vac, 50/60 Hz
,C3C for Isolated 9-32 Vdc
,C3E for Isolated 26-56 Vdc
(4) ADD-ON OPTIONS
  Select
Nothing(leave field blank) for Red LED display
,G for Green LEDs for display
(5) ADD-ON OPTIONS
  Select
Nothing(leave field blank) for No Options
,BL for Lens without Newport logo
 NOTE: All combinations may not be valid, check spec sheet for valid part numbers.


   
Keywords: 3 1/2 Digit Thermocouple Meters